LMG611
|
The Most Precise in its Class
- Outstanding accuracy of 0.015% of measured value + 0.01% of range
- Full dynamic range of 500 µA to 32 A / 3 mV to 1000 V per channel available in single instrument
- Analog bandwidth DC up to 10 MHz
- Simultaneous measurement of narrow- and broadband values through innovative DualPath architecture
- Harmonics and interharmonics up to 2000. order, as required by EN61000-4-7
- With optional I/O card speed/torque inputs freely configurable for all signal types (analogue, frequency as RS422, TTL or HTL) via menu
- Flexible scripting tool for custom applications
- Simultaneous measurement of V, I, P values and harmonics, presentation in tabular or graphical form
- Flicker measurement, interactions between grid and appliance according to EN61000-4-15
- Large and clearly arranged touchscreen GUI for intuitive handling
- Internal storage of long-term measurements even with shortest cycle-time thanks to integrated mass storage device
- Excellent connectivity via USB2.0, Gbit-LAN, RS232, and DVI
- Convenient "plug-and-measure" through automated sensor detection
- Calibration certificate included free-of-charge in first delivery
- 24 months warranty
LMG611-A for high-precision broadband measurements
High-Precision Channel for power measurements with DualPath functionality, bandwidth DC-10 MHz
best accuracy 0.015% + 0.01%
Measuring inputs and ranges:
1. Voltage directly (V): 3/6/12,5/25/60/130/250/400/600/1000Vtrms; 3200 Vpk
2. Voltage sensor input (mV): 30/60/120/250/500 (V): 1/2/4Vtrms; 8 Vpk
3. Current directly (mA): 5/10/20/40/80/150/300/600 and (A):1.2/2.5/5/10/20/32 Atrms; 120 Apk
4. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4Vtrms; 12.5 Vpk
LMG611-B for general purpose measurements
Multi-purpose measurement channel with DualPath functionality, bandwidth DC - 500 kHz
best accuracy 0.05% + 0.02%
Measuring inputs and ranges
1. Voltage directly (V): 3/6/12.5/25/60/130/250/400/600/1000Vtrms; 3200 Vpk
2. Current directly (mA): 5/10/20/40/80/150/300/600 and (A):1.2/2.5/5/10/20/32 Atrms; 120 Apk
3. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4Vtrms; 12.5 Vpk
LMG611-C for high-precision 50/60 Hz measurements
High-precision measurement channel for 50 Hz applications, bandwidth DC - 10 kHz,
best accuracy 0.03% + 0.01%.
Measuring inputs and ranges:
1. Voltage directly (V): 3/6/12,5/25/60/130/250/400/600/1000Vtrms; 3200 Vpk
2. Current directly (mA): 5/10/20/40/80/150/300/600 and (A):1.2/2.5/5/10/20/32 Atrms; 120 Apk
3. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4Vtrms; 12.5 Vpk
|
|
|